Wednesday, July 10, 2013
Mock MRI simulation for pediatric MRI.....
The ‘mock MRI’ procedure involves children undergoing a simulated scan with the assistance of a play therapist prior to the actual diagnostic scan. It acts as both a screening tool, to assist in identifying children who are likely to be able to comply with the MRI procedure,and also helps to prepare these children,by familiarizing them with the environment, sounds, and equipment, while teaching them skills (such as breathing,relaxation, or distraction) to cope with the actual procedure . Use of the ‘mock’ magnet has led to a marked reduction in the numbers of patients who have required anesthetic and reduced the time required for the diagnostic scan......
Saturday, July 6, 2013
Pediatric MRI..............
Longer TR and TE times are generally required when imaging very young children.the high water content in neonatal brains, coupled with the lack of fatty myelin results in the reduction in CNR (contrast to noise ratio) and grey/white matter differentiation.restore pulses on T2W imaging can improve CSF contrast and allow shorter TR times to reduce scan time......
Tuesday, July 2, 2013
EMERGENCIES IN MR ENVIRONMENT
EMERGENCIES IN MR ENVIRONMENT
3 categories:
- Quench
- Magnetic field emergencies
- Patient emergencies
QUENCHING:
- Sudden loss of absolute zero of temperature in the magnet coils
- Coils become resistive
- Rapid escape of helium from cryogen bath
- Accidental
quenching
- Manual quenching
- In case of
emergency (if a person pinned to a magnet
by a large ferromagnetic object)
- can cause severe, irreparable damage to the superconducting
coils
System requirements for quenching:
- Helium venting equipment
- Oxygen monitor
- Alarms when oxygen levels fall below certain level
Steps in case of a quench:
- Do not panic
- Turn on exhaust (if not automatically turned on)
- Open the door between operator room and hallway
- Enter the scan room—undock table—help patient exit
- Evacuate area and return only when oxygen levels are normal.
- IF HELIUM IS VENTING INTO THE ROOM THE SCAN ROOM MAY NOT OPEN…..
In that case:
- Try opening the door several times, but
-- If not opened in 45 seconds
- Break window to
scan room to relieve
Pressure
- Enter the scan
room
- Evacuate
patient
MAGNETIC FIELD
EMERGENCIES:
- If somebody pinned against the magnet or in other magnetic-field-related emergency
QUENCH THE MAGNET
But
It results in several
days’ downtime
So
Do not press QUENCH BUTTON unless true emergency
- Do not test the button, leave to qualified service personnel
PATIENT
EMERGENCY:
- Conditions
- Seizure or
claustrophobia
- Greater potential
of cardiac arrest
- Unconscious,
heavily sedated or confused patient
- Monitor these patients closely as usually direct observation
of the patient not possible due to partial obscuration by magnet enclosure
- Emergency personnel either remains with the patient or on
stand-by alert
Procedure to be followed in case of patient emergency:
- Hit emergency stop button
- Evacuate patient to the emergency medical treatment area
outside exclusion zone
- Follow emergency hospital protocol
BS(RT)
Subscribe to: